LOG-file for Pluvio2 meaning Pluvio 200 inside wind fence, connected to port COM4 20140115 5 l of anti-freeze is added. Anti-freeze equals now 10l. Power-cuts and usb-cuts are tested.File name convention changed, data folder is now Pluvio200_02 (Pluvio200_01). Annakaisa von Lerber 20140129 After last time anti-freeze filling one screw of the cover was loose and this caused trembling and data is unusable from 15 Jan till today. This is fixed today. Annakaisa von Lerber 2014-01-31 13:55 UTC installations finished on the platform. Matti Leskinen 2014-05-12 18:15 - 18:58 UTC Pluvio2 200 bucket is emptied and switched to outside the wind fence. The file is called now 200 and _os_.Annakaisa 2014-05-19 the base tubes of the both pluvios were switched so that their tops would match the wind shield levels. During the process the horizontality of the scale plate was checked as well. Matti Leskinen 2014-06-11 the Pluvio200_os was not working after thunderstorm, but and testing was performed between the instruments. For some reason it started to work again. -Annakaisa 2014-06-30 PC clock set manually at 11:58:30 UTC, the clock was late about 2:50 minutes. Matti Leskinen 2014-08-25 Switched the weighing scales of the two pluviometers, and changed the instrument labels. After the changes Pluvio400_i0 and Pluvio200_o0 are used for the two systems. Matti Leskinen 2014-09-29 Pluviometers were switched back to their original positions. Pluvio200_02 on the platform inside the fence, and Pluvio400_01 inside the Alter shield on the field. During the switch insects and other macroscopic objects were taken away from the liquids. Matti Leskinen 2014-11-05 The liquid is changed to more stronger, the ratio is now 2/2 with the anti-freeze. About 10 l was put. -Annakaisa v. Lerber 2014-11-17 Testing if the Pluvio software could be upgraded, but no driver found in laptop. During the process it became clear that the recording was switched for Pluvio400 and Pluvio200. The recording was switched to use "COM5" port from the right serial input cable beginning 14:25 UTC. Matti Leskinen 2014-11-18 The problem with wrong instrument in the files was being corrected and data labels were updated, PluvioErr folder was created to store the original files. Matti Leskinen 2014-11-24 Ott Pluvio2 system software upgrade V.1.31.0, Recording break around 10:15-12:30 UTC. Matti Leskinen 2014-12-03 Time check, Wed Dec 3 10:15:45 UTC 2014: Pluvio 10:14:00. Time adjusted manually 11:05:45 UTC. Matti Leskinen 2014-12-15 Short logger break to check the voltages sent to the Pluvios. Found 24V OK for both. Changed W2TIME update interval from 360000 to 7200. Matti Leskinen 2015-01-04 Record stopped around 14:20 UTC, restarted 2015-01-05 07:20 UTC. Matti Leskinen 2015-01-07 Added antifreeze into the container,around 12:10 UTC: 815mm to 893mm. Matti Leskinen 2015-01-14 Recording had stopped in the morning 04 UTC. Restarted 2015-01-16 08:50 UTC. Time difference to reference found -42 s, time changed manually. Matti Leskinen 2015-01-26 Pluviometer bucket was emptied after 12 UTC. New antifreeze solution was added, and this meant a step change in recorded weight to about 450 mm. At the same time the serial-USB was changed to use "COM7" port. Annakaisa von Lerber, Matti Leskinen 2015-01-27 15:20 UTC circa 1 dl of sunflower oil is added to prevent condensation. Annakaisa von Lerber, Matti Leskinen 2015-02-10 Time difference -43 s, clock adjusted manually 14:56:12 UTC. Matti Leskinen 2015-02-13 Swapped the names of the logger scripts, Pluvio200 is recorded by pluvio_logger2, and at the same time changed the logger poll time 1 second later than the other one. Matti Leskinen 2015-02-17 Recording had stopped 04:50 UTC. Restarted logger 15:36, OK. Matti Leskinen 2015-03-20 Recording had stopped on the 19th 22:03 UTC, Restarted logger 10:30 UTC, OK. Matti Leskinen 2015-04-13 Maintenance operation 11-12 UTC. The input voltage of the system was fixed to be nominal 24 V, for a while the voltage level was about 8 V according to the records, but this was not serviced in the winter while the data seemed to be OK. The problem was located to the cable's red wire, and fixed by using other wire pair. Matti Leskinen 2015-04-27 Checking the bucket position at about 09:30 UTC, because of suspecting noise level increase since the last maintenance operation. The shield seemed to be a bit more easy to attach over the bucket after the check. Remained to be seen if this had any effect on the noise in windy weather. The PC clock was set to match the real time according to the site server at 14:13:30 UTC, the PC clock was about 2 minutes lagging at the time. Matti Leskinen 2015-05-29 PC time behind 54 seconds, time set 07:49:40 UTC. Matti Leskinen 2015-06-17 Emptied the bucket and leveled the instrument. The oil had created a plastic foil, and was hard to remove. Not a good idea. Annakaisa 2015-08-11 Recording had stopped around 09:30 UTC, restarted logger at about 21 UTC. PC clock was running behind about 2.5 minutes, and the clock was set at 21:05:55 UTC. Matti Leskinen 2015-08-17 Recording had stopped on the 16th after 09:21 UTC, restarted logger 14:37 UTC. Matti Leskinen 2015-08-25 Recording had stopped 00:45 UTC, restarted 13:28 UTC. Matti Leskinen 2015-08-31 Recording had stopped 02:13 UTC, restarted 10:34 UTC. Matti Leskinen 2015-09-07 Pluviometer recording was moved to "Hotplate" laptop, starting at about 09:10 UTC. Matti Leskinen 2015-10-31 Pluviometer recording had stopped on the 30th, restarted 11:53 UTC. Matti Leskinen 2015-11-25 Restarted 11:55 UTC. Since 20th a lot of power cuts and restarts. Matti Leskinen 2016-01-04 Antifreeze liquid was added 13 UTC. Matti Leskinen 2016-02-12 Recording had stopped on the 11th at 15:02 UTC. Logger restarted 12:06 UTC. Matti Leskinen 2016-04-01 Recording had stopped on March 31st 14 UTC, restarted 07:12 UTC. Matti Leskinen 2016-04-05 Recording had stopped on the 4th 14:41 UTC, restarted logger 06:43 UTC. Matti Leskinen 2016-04-10 Recording had stopped 01:23 UTC, restarted logger 06:26 UTC. Matti Leskinen 2016-04-24 Recording had stopped 14:10 UTC, restarted 18:07 UTC. Matti Leskinen 2016-06-01 Mistakenly shutting down the system in the morning while figuring out Pluvio400 problem. Start of the system at about 13 UTC by Janne Levula. Matti Leskinen